Oscillations
The oscillation therapy feature delivers an oscillatory therapy based on frequency and amplitude settings. Use of the oscillation feature enhances mobilization and improves bronchial drainage.
Inspiratory Trigger
An important characteristic of the system is its ability to trigger on the patient’s inhale to help synchronize the therapy with the patient. The inspiratory trigger feature is available when the device is in Auto mode. The pressure delivery sequence is synchronized with the patient’s effort to inhale.
Inspiratory Flow
While the expiratory switching is desired to be fast and at high flow to simulate cough, the inspiratory flow may be controlled to a comfortable level by
selecting the flow between low, medium and high.
Positive Airway Pressure on the Pause Phase
Provides the option to keep airways open in-between the exhalation and inhalation phases of the cycle cough. This feature is designed to make the therapy more comfortable for the patient.
Included:
• BiWaze® Cough System
• Single-Path Breathing Tube
• Patient Port Adapter for 22mm Filter
• Adult Medium Face Mask
• Battery
• Power Cord
• Carrying Bag
• User Manual
Specifications:
Operating temperature | 40° to 95°F (5C to 35C) |
Relative humidity | 15 to 95% (non condensing) |
Size | 27 cm L x 23.5 cm W x 9.0 cm H (10.5″ L x 9.2″ W x 3.5″ H) |
Weight | 9lbs (4kg) including internal battery |
Presets | Up to 10 |
Modes | Automatic and Manual |
Maximum positive pressure | 70 cm H20 |
Maximum negative pressure | -70 cm H20 |
Maximum pause pressure | 15 cm H20 |
Inspiratory time | 0 – 5 seconds (auto mode only) |
Expiratory Time | 0 – 5 seconds (auto mode only) |
Pause Time | 0 – 5 seconds (auto mode only) |
Inspiratory Flow | Low, Medium, High |
Inspiratory trigger sensitivity | 1 – 10 |
Oscillation frequency | 5 – 20 Hz |
Amplitude | 1 – 5 |
Warranty |
1 Year on Device 90 Days on Battery 30 Days on Carry Bag 90 Days on Foot Pedal |