Description
The Thermovent T2 is s sterile, single-use heat and moisture exchanger that provides effective humidification for spontaneously breathing patients whose upper airways are bypassed by a tracheal or tracheostomy tube.
These devices are used to:
- Treat adult or pediatric patients with a tidal volume of more than 70 mL
- Reduce heat loss through the bronchopulmonary tree
- Help to prevent thickened secretions and changes in lung function
- Supply additional oxygen when required
- Facilitate the removal of secretions
- Preserve humidity and minimize damage to tracheal epithelial cells
Product Benefits:
- Simple opening port allows quick and easy access for a suction catheter to aid the removal of secretions.
- Integrated oxygen port that can provide up to 60% oxygen concentrations to the patient via the simplest of connections.
- Original Portex™ Thermovent T paper media provides high-level performance heat and moisture exchange to patient.
- Smooth, rounded edges aid patient comfort.
Oxygen Flow Rates/Inspired Oxygen Concentration | ||||||||||
02L/min | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 |
02% | 27 | 33 | 37 | 42 | 45 | 47 | 53 | 54 | 57 | 60 |
Moisture Output: 25mgL-1 H2O @ 15 breath min-1 Tidal Volume 500mL
Resistance to Flow |
0.3 hPa (cmH2O) at gas flow rate 0.5L sec-1 (30L min-1) [0.6cm H2O L-1 sec-1] |
0.7 hPa (cmH2O) at gas flow rate 1.0L sec-1 (60L min-1) [0.7cm H2O L-1 sec-1 |
1.2 hPa (cmH2O) at gas flow rate 1.5L sec-1 (90L min-1) [0.9cm H2O L-1 sec-1] |
Dead Space: 11mL
Weight: 7.5g
Inspired Oxygen Concentration (%): 15 breath/min
Tidal Volume: 500mL
Features